GRAPHTEC Multi-channel Data Logger GL260

Layanan Baru: G-REMOTE

Fitur Baru G-REMOTE

  • Alarm History → Menyimpan dan menampilkan riwayat alarm secara detail.

  • Improved Inter-CH Calculation → Perhitungan antar-channel lebih akurat dan cepat.

  • Memory Loop Function → Perekaman data berkelanjutan tanpa kehilangan data penting.

Dengan adanya ketiga fungsi ini, kemudahan penggunaan (usability) meningkat secara signifikan.

Kelebihan Utama

  • Mendukung semua channel isolated dan multi-function input.

  • Dengan mempersempit jumlah channel, interval sampling hingga 10ms dapat dicapai untuk perekaman data cepat.

  • Unit WLAN (opsional) memungkinkan pengukuran nirkabel menggunakan PC dan pemantauan jarak jauh (remote monitoring).

Spesifikasi GL260 Data Logger

ItemDescription
Number of analog input channels10 channels
External input/outputInput: Trigger/Sampling (1 ch), Logic/Pulse (4 ch)
• Voltage range: Up to 24V (common ground)
• Signal type: Voltage, Open collector, Relay contact
• Threshold: Approx. +2.5V (Hysteresis: 2.5V–3V)Output: Alarm (4 ch)
• Output signal: Open collector (pull-up to 5V by 10kΩ resistor)
• Max. rating: 30V / 0.5A / 0.2W
Sampling interval10 ms to 1 hour (10ms–50ms for voltage only; minimum interval depends on number of channels), External signal
Time scale of waveform display1 sec to 24 hours/division
Trigger / Alarm functionTrigger action: Start/Stop capture by trigger
Repeat action: Off / On (auto rearmed)
Trigger source: Off, Measured signal, Alarm, External, Clock, Week, Time
Condition setting: OR/AND combination
Analog signal: Rising, Falling, Window-in, Window-out
Logic signal: Pattern (high/low combination)
Pulse signal: Rising, Falling, Window-in, Window-out
Pulse input functionMeasurement modes: Rotation count (RPM), Accumulating count, Instant count
• Max. frequency: 50kHz
• Max. count: 50kC/sampling (16-bit counter)
Interface to PCUSB (Hi-speed), WLAN (optional B-568, available in limited regions)
Storage deviceInternal: Built-in 8GB Flash
External: SD card slot (SDHC up to 32GB)
Saved contents: Captured data, Settings, Screen copy
Display4.3-inch TFT color LCD (WQVGA, 480×272 dots)
Operating environment0–45°C, 5–85% RH (non-condensed)
With battery pack: 0–40°C (operation), 15–35°C (charging)
Power sourceAC adapter: 100–240V AC, 50/60Hz (1 included)
DC power: 8.5–24V DC (optional cable B-514)
Battery pack: Optional B-573, 7.2V DC, 2875mAh
Power consumptionMax. 36 VA (with LCD ON, AC adapter use, and charging)
External dimensions (W×D×H)Approx. 188 × 117 × 42 mm (excluding projections)
WeightApprox. 500 g (excluding AC adapter & battery pack)

Spesifikasi Teknis Input Terminal GL260

ItemDescription
Type of input terminalM3 screw type terminals (dengan rectangular flat washer)
Input methodPhoto MOS relay scanning system
Semua channel isolated, balanced input
Measurement rangeVoltage: 20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20, 50, 100 V; 1–5 V F.S.
Thermocouple: K, J, E, T, R, S, B, N, C (W: WRe5-26)
Humidity: 0–100% RH (konversi scaling 0–1 V) dengan sensor kelembapan opsional B-530
FilterOff, 2, 5, 10, 20, 40 (moving average sesuai jumlah yang dipilih)
A/D converterSigma-Delta type, 16-bit (resolusi efektif: 1/40000 dari full measuring range)
Maximum input voltageAntar terminal (+/–): 20 mV–1 V range = 60 Vp-p
2 V–100 V range = 110 Vp-p
Antar input terminal: 60 Vp-p
Input terminal – GND: 60 Vp-p

 

Spesifikasi GL28-APS Application Software

ItemDescription
Model nameGL28-APS
Supported OSWindows 11 (64-bit) / Windows 10 (64-bit / 32-bit)
FunctionsKontrol perangkat seri GL, real-time data capture, konversi format data
Number of groupsHingga 4 grup
Number of CHs per groupMaksimal sesuai jumlah modul yang terhubung
Maximum number of channelsHingga 1000 channel
Settings controlPengaturan amplifier, capture, trigger/alarm, report, dan lainnya
Capturing dataReal-time data dalam format CSV & GBD Binary
Data tersimpan di internal memory atau SD card (CSV, GBD Binary)
DisplayTampilan analog waveform, logic waveform, pulse waveform, digital values
Display modesY-T View, Digital View, X-Y View antar cursor (hanya saat replay)
File conversionAntara cursor atau seluruh data
Monitor functionsAlarm monitor dapat mengirim email ke alamat yang ditentukan
Statistic / HistoryMenampilkan nilai maksimum, minimum, dan rata-rata selama pengukuran
Report functionMembuat laporan harian atau bulanan secara otomatis

Dimensi

GL260 External Dimensions

Scale : mm
Tolerance : +/- 3 mm

GL260 Terminals

Scale : mm
Tolerance : +/- 0.5 mm